Raman Specifications
| Spectrometer and Detectors | |||||||
| Spectrometers | MicroHR | iHR320 | iHR550 | ||||
| Excitation Lasers | 532 nm, 633 nm, 785 nm | ||||||
| Spectral Range (cm-1) – Free space | 80 – 9500 (532 nm), 80 – 6500 (633 nm), 80 – 3400 (785 nm) | ||||||
| Spectral Range (cm-1) – Fiber | 150 – 9500 (532 nm), 150 – 6500 (633 nm), 150 – 3400 (785 nm) | ||||||
| Recommended Gratings | 1800 g/mm, 1200 g/mm, 600 g/mm | ||||||
|
Spectral Resolution1 (cm-1/ pixel) |
405 nm | 5.12 | 2.34 | 1.37 | |||
| 532 nm | 2.53 | 1.22 | 0.73 | ||||
| 633 nm | 1.54 | 0.78 | 0.48 | ||||
| 785 nm | 0.70 | 0.40 | 0.26 | ||||
| Microscope | |||||||
|
Microscope Objectives |
Magnification | 10X | 50X | 100X | |||
| Spot Size
(fiber-coupled) |
< 50 μm | < 12 μm | < 6 μm | ||||
| Spot Size
(free space-coupled) |
< 10 μm | < 5 μm | < 2 μm | ||||
| Sample Stage | XYZ (Manual and motorized options available) – 75 x 50 mm; 100 x 100 mm; 150 x 150 mm; 300 x 300 mm | ||||||
| Vision Camera | Software controlled vision camera included | ||||||
1 For 1800 g/mm grating and 26 μm pixel CCD
Photoluminescence Specifications
| Spectrometer and Detectors | ||||||
| Spectrometers | MicroHR | iHR320 | iHR550 | |||
| Excitation Lasers1 | 266 nm, 325 nm, 405 nm, 532 nm, 633 nm, 785 nm, 980 nm, 1064 nm | |||||
| Spectral Range (nm)2 | 250 nm – 2200 nm | |||||
| Recommended Gratings3 | 1800 g/mm, 600 g/mm, 300 g/mm | |||||
| Spectral Resolution4 (nm) | 0.39 | 0.18 | 0.1 | |||
| Microscope | ||||||
|
Microscope Objectives5 |
Magnification | 10X | 50X | 100X | ||
| Spot Size
(fiber coupled) |
< 50 μm | < 12 μm | < 6 μm | |||
| Spot Size
(free space coupled) |
< 10 μm | < 5 μm | < 2 μm | |||
| Sample Stage | XYZ (Manual and motorized options available) – 75 x 50 mm; 100 x 100 mm; 150 x 150 mm; 300 x 300 mm | |||||
| Vision Camera | Software controlled vision camera included | |||||
1Other laser wavelengths available on request
2 Extension into mid IR available on request
3 Other gratings available on request
4 Based on 1200 g/mm grating at 500 nm and a 26 μm pixel CCD
5 Reflective objectives may be recommended if working in the UV or using multiple sources that cover a broad spectral range
Time-resolved Photoluminescence (Lifetime) Specifications
| Spectrometer and Detectors | ||||||
| Spectrometers | MicroHR | iHR320 | iHR550 | |||
|
Excitation Lasers |
Discrete pulsed sources available from 260 nm to 1060 nm | |||||
| Tunable pulsed white laser from 400 nm | ||||||
| Spectral Range (nm) | 250 nm – 1700 nm | |||||
| Recommended Gratings1 | 1200 g/mm, 600 g/mm, 300 g/mm | |||||
| Spectral Resolution2 (nm) | 0.39 | 0.18 | 0.1 | |||
| Measurable Lifetime Range3 | 25 ps to seconds | |||||
| Microscope | ||||||
|
Microscope Objectives4 |
Magnification | 10X | 50X | 100X | ||
| Spot Size
(fiber-coupled) |
< 50 μm | < 12 μm | < 6 μm | |||
| Spot Size
(free space-coupled) |
< 10 μm | < 5 μm | < 2 μm | |||
| Sample Stage | XYZ (Manual and motorized options available) – 75 x 50 mm; 100 x 100 mm; 150 x 150 mm; 300 x 300 mm | |||||
| Vision Camera | Software controlled vision camera included | |||||
| Additional software5 | EzTime (or EzTime Image, if doing FLIM) | |||||
1 Other gratings available on request
2 Based on 1200 g/mm grating at 500 nm and a 26 μm pixel CCD
3 Electronics and detectors for lifetimes shorter that 25 ps available on request
4 Reflective objectives may be recommended if working in the UV or using multiple sources that cover a broad spectral range
5 Both EzTime and EzTime Image can be installed on the same computer as LabSpec 6
Reflectance and Transmittance Specifications
| Spectrometer and Detectors | ||||||
| Spectrometers | MicroHR | iHR320 | iHR550 | |||
|
Excitation Sources |
Xenon Lamp (200 nm – 2200 nm) | |||||
| Tungsten Halogen Lamp ( > 350 nm) | ||||||
| Supercontinuum Laser ( > 400 nm) | ||||||
| Spectral Range (nm)1 | 250 nm – 2200 nm | |||||
| Recommended Gratings2 | 1200 g/mm, 600 g/mm, 15 0 g/mm | |||||
| Spectral Resolution3 (nm) | 0.39 | 0.18 | 0.1 | |||
| Microscope | ||||||
|
Microscope Objectives4 |
Magnification | 10X | 50X | 100X | ||
| Spot Size
(fiber-coupled) |
< 50 μm | < 12 μm | < 6 μm | |||
| Spot Size
(free space-coupled) |
< 10 μm | < 6 μm | < 2 μm | |||
| Sample Stage | XYZ (Manual and motorized options available) – 75 x 50 mm; 100 x 100 mm; 150 x 150 mm; 300 x 300 mm | |||||
| Vision Camera | Software controlled vision camera included | |||||
1 Extension into mid-IR available on request
2 Other gratings available on request
3 Based on 1200 g/mm grating at 500nm and a 26 μm pixel CCD
4 Reflective objectives may be recommended if working in the UV or using multiple sources that cover a broad spectral range
Electroluminescence Specifications
| Spectrometer and Detectors | ||||||
| Spectrometers | MicroHR | iHR320 | iHR550 | |||
| Excitation Source1 | Keithley Source Meter (2400 series) | |||||
| Spectral Range (nm) | 250 nm – 14 μm | |||||
| Recommended Gratings2 | 1200 g/mm, 600 g/mm, 150 g/mm | |||||
| Spectral Resolution3 (nm) | 0.39 | 0.18 | 0.1 | |||
| Microscope | ||||||
| Microscope Objectives4 | Magnification | 10X | 50X | 100X | ||
| Sample Stage | XYZ (Manual and motorized options available) – 75 x 50 mm; 100 x 100 mm; 150 x 150 mm | |||||
| Vision Camera | Software controlled vision camera included | |||||
1 Other current voltage sources available on request
2 Other gratings available on request
3 Based on 1200 g/mm grating at 500 nm and a 26 μm pixel CCD
4 Reflective objectives may be recommended if working in the UV, or using multiple sources that cover a broad spectral range
Photocurrent Specifications
| Spectrometer and Detectors | ||||||
| Spectrometers | MicroHR | iHR320 | iHR550 | |||
|
Excitation Sources |
Lasers: 266 nm – 1064 nm | |||||
| Xenon Lamp (200 nm – 2200 nm) – Tunable option available | ||||||
| Tungsten Halogen Lamp ( > 350 nm) – Tunable option available | ||||||
| Supercontinuum Laser ( > 400 nm) – Tunable option available | ||||||
| Photocurrent Measurement1 | Keithley Source Meter (2400 series) | |||||
| Recommended Gratings2 | 1200 g/mm, 600 g/mm, 150 g/mm | |||||
| Spectral Resolution3 (nm) | 0.39 | 0.18 | 0.1 | |||
| Microscope | ||||||
| Microscope Objectives4 | Magnification | 10X | 50X | 100X | ||
| Sample Stage | XYZ (Manual and motorized options available) – 75 x 50 mm; 100 x 100 mm; 150 x 150 mm; 300 x 300 mm | |||||
| Vision Camera | Software controlled vision camera included | |||||
1 Extension into mid-IR available on request
2 Other gratings available on request
3 Based on 1200 g/mm grating at 500 nm and a 26 μm pixel CCD
4 Other objectives may be recommended if working in the UV, or using multiple sources that cover a broad spectral range
Darkfield Scattering Specifications
| Spectrometer and Detectors | ||||||
| Spectrometers | MicroHR | iHR320 | iHR550 | |||
|
Excitation Sources |
Xenon Lamp (200 nm – 2200 nm) | |||||
| Tungsten Halogen Lamp ( > 350 nm) | ||||||
| Supercontinuum Laser ( > 400 nm) | ||||||
| Spectral Range (nm)1 | 200 nm – 2200 nm | |||||
| Recommended Gratings2 | 1200 g/mm, 600 g/mm, 150 g/mm | |||||
| Spectral Resolution3 (nm) | 0.39 | 0.18 | 0.1 | |||
| Microscope | ||||||
| Microscope Objectives4 | Magnification | 10X | 50X | 100X | ||
| Sample Stage | XYZ (Manual and motorized options available) – 75 x 50 mm; 100 x 100 mm; 150 x 150 mm; 300 x 300 mm | |||||
| Vision Camera | Software controlled vision camera included | |||||